Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 523/723/523K/723K pump is intended for use in the electromagnetic environment
specified below. The customer or user of the Paradigm 523/723/523K/723K pump system should
make sure that it is used in such an environment.
Immunity
Test IEC 60601
Level
Compliance
Level Electromagnetic Environment Guidance
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption, and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcast cannot be
predicted theoretically with accuracy. To access the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Paradigm 523/723/523K/723K pump is used exceeds the applicable RF compliance
level above, the Paradigm 523/723/523K/723K pump should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Paradigm 523/723/523K/723K pump.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m.
Pump specifications202