Medtronic Guardian Real-time Betriebsanweisung Seite 122

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 184
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 121
Warning: If you see any body fluid in the connector opening, you must discard the transmitter.
Because the transmitter contains a battery, do not discard in a bio-waste container. Instead,
continue to clean and disinfect the transmitter, and then discard according to local
regulations for battery disposal (non-incineration).
connector opening
connector pins
housing
3 Attach the tester to the transmitter.
4 If optional occlusive dressing was used and there is adhesive residue on the transmitter, follow the
instructions in Removing adhesive residue, on page 110.
5 Rinse the transmitter under running room temperature tap water for at least one minute and until any
visible debris is gone.
6 Prepare an enzymatic solution using two (2) ounces of ENZOL Enzymatic Detergent per gallon of
water.
7 With the tester still attached, fully submerge the transmitter in the enzymatic solution for one
minute.
System maintenance 107
Chapter 10
Seitenansicht 121
1 2 ... 117 118 119 120 121 122 123 124 125 126 127 ... 183 184

Kommentare zu diesen Handbüchern

Keine Kommentare