Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Guardian REAL-Time system (consisting of the CSS7100/CSS7100K Receiver and MMT-7703
transmitter) is intended for use in the electromagnetic environment specified below. The customer
or the user of the Guardian REAL-Time system should make sure that it is used in such an
environment.
Immunity Test IEC 60601Test Level Compliance
Level
Electromagnetic Environment
- Guidance
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Guardian REAL-Time is intended for use in the electromagnetic environment specified below.
The customer or user of the Guardian REAL-Time system should make sure that it is used in such an
environment.
Immunity
Test
IEC 60601
Level
Compliance
Level
Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
10Vrms
150kHz to
80MHz
Not applicable
Portable and mobile RF communications equipment
should be used no closer to any part of the
Guardian REAL-Time, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Not applicable
System specifications 125