Medtronic Paradigm Link Betriebsanweisung Seite 187

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 262
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 186
Alarm conditions
Alarms put the pump in Attention mode.
A (ALARM)
This alarm shows a letter A followed by two numbers. A-alarms cause all insulin delivery to stop. Your pump
settings are retained. If this alarm repeats often, call the HelpLine for assistance.
AUTO OFF
Alerts you that no buttons were pressed during the time limit you set for the AUTO OFF DURATION feature,
and so insulin delivery has been stopped.
BATT OUT LIMIT
Occurs if the battery has been out of the pump for more than five minutes. Verify that the pump time and
date are correct. If not correct, go to the UTILITIES MENU and reset the date and time.
BOLUS STOPPED
If this alarm occurs, it is very important to check bolus history to review how much of the bolus was actually
delivered. If necessary, reprogram a bolus with the amount not yet delivered.
BUTTON ERROR
Occurs if a button has been continually pressed for more than three minutes.
CHECK SETTINGS
When this alarm is active, you should check and/or reprogram your pump settings, including the time/date.
E (ERROR)
An E-error alarm will show the letter E followed by two numbers. E-alarms cause all insulin delivery to stop,
the pump resets, and all your settings are cleared. After receiving this alarm, note the error number then
call your local Medtronic Diabetes representative for assistance.
Troubleshooting and alarms
168
Seitenansicht 186
1 2 ... 182 183 184 185 186 187 188 189 190 191 192 ... 261 262

Kommentare zu diesen Handbüchern

Keine Kommentare