Medtronic Paradigm Link Betriebsanweisung Seite 83

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 262
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 82
1
If you are using the pump for the first time, remove the red shipping cap from the reservoir compartment.
WARNING: Do not insert the reservoir in the pump if you did not rewind. Doing so could result in
inaccurate insulin delivery.
Never insert the reservoir into the pump while the tubing is connected to your body. Doing so could
result in an accidental infusion of insulin.
2
Insert the reservoir into the top of the pump case.
ESC
3
Turn the tubing connector approximately 1/2 -turn clockwise until the connector is seated. The tubing
connector should be aligned horizontally with the pump case as shown here.
tubing connector
4
Attach the activity guard, if desired.
5
You must now do a manual prime as described in the next section.
Starting on insulin
64
Seitenansicht 82
1 2 ... 78 79 80 81 82 83 84 85 86 87 88 ... 261 262

Kommentare zu diesen Handbüchern

Keine Kommentare