Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 522/522K/722/722K pump system is intended for use in the electromagnetic environ-
ment specified below. The customer or user of the Paradigm 522/522K/722/722K pump system should
make sure that it is used in such an environment.
Immunity
Test
IEC 60601
Level
Compliance
Level
Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
10Vrms
150kHz to
80MHz
10V/m
80MHz to
6.0GHz
Not applicable
10 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the Paradigm
522/722/522K/722K pump, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Not applicable
d = 0.35
P
80 MHz to 800 MHz
d = 0.70
P
800 MHz to 6.0 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey
a
, should be
less than the compliance level in each frequency
range
b
.
Pump specifications
201
Chapter 15