Entering meter BG for calibration
Sensor calibration is necessary for optimal glucose sensor performance. Only BG entries in the range of 40–
400 mg/dL (2.2–22 mmol/L) are accepted for sensor calibration.
Use any of the following methods for your sensor calibrations, including the first calibration after the
initialization. You can enter BG values manually or through your linked meter.
To enter BG values manually:
1 You can enter a meter BG value manually in the Bolus Wizard.
a.
Press the
button on your pump, or use the MAIN MENU to go to the ENTER BG screen:
Main > Bolus > Use Bolus Wizard
b. Use the up and down arrow buttons on your pump to enter your meter BG. Press ACT.
c. Select Yes to calibrate, or No to cancel, when the following screen displays: BG to update Sensor?
2
You can enter a meter BG value manually through the Sensor menu.
a. Follow this path:
Main > Sensor > Calibrate
b. Enter the BG value, and then press ACT. Your sensor is now calibrated with the BG value you
entered.
• If the Bolus Wizard feature is off, the SENSOR MENU screen appears.
• If the Bolus Wizard is on and the saved BG is below your BG Target range, the Cal Saved screen
appears. This screen indicates that your saved BG is below your BG Target set in the Bolus Wizard
feature, and the pump beeps three times. See Setting the BG Targets, on page 71 for more
information.
• If the Bolus Wizard is on, the saved BG is above your BG Target range, and the calculated
correction estimate is more than the preset scroll rate increment, the Cal Saved screen appears.
This screen indicates that your saved BG is above your BG Target set in the Bolus Wizard feature,
and the pump beeps three times. Go to the ENTER BG screen of the Bolus Wizard feature to enter
a correction bolus. See Using the Bolus Wizard feature section for more information.
• If the information is not correct, select Cancel, then press ACT. The SENSOR MENU screen
reappears with Calibrate selected. Repeat the procedure and enter the correct information.
3
You can enter a meter BG value manually through the Capture Event menu. Follow this path:
Main > Capture Event > Enter BG
Sensor features116