
Example 2: Block
Oscar is an elderly man with diabetes who needs assistance with all of his daily living activities. He
needs his caregiver or family member to help him with his pump as well. To be sure that Oscar does not
change any pump settings, his family programmed his Paradigm pump with the Block feature turned on.
They use the Remote Control to give him his boluses when he needs them.
Lock keypad feature
Lock keypad prevents accidental pump keypad presses. You can only press the to view the STATUS
screen, and the
button to turn on the backlight. The remote control can be used to give a bolus or put
the pump into Suspend.
A locked keypad is automatically unlocked during the following:
• Battery insertion
•Alarms
•Alerts
Locking the keypad
1
Go to the UTILITIES MENU.
2 Select Lock Keypad, then press ACT.
3 Press ACT again to lock the keypad. The KEYPAD LOCKED screen appears with instructions on how to
unlock the keypad.
Unlocking the keypad
Press the button and the button at the same time. Keypad Unlocked screen appears.
Selftest
Selftest is a safety utility that allows you to check if your pump is operating properly. This self-diagnostic
feature can be used for maintenance or to check your pump if it operates unusually. During selftest, your
pump will automatically run internal tests, including a check for proper operation of the beep and vibrate
modes. The selftest is additional to the routine tests that run independently while the pump operates.
Utilities 141
Chapter 9
Kommentare zu diesen Handbüchern