Square Wave bolus
Delivers bolus insulin over a duration of 30 minutes up to eight hours (limited by the Max bolus setting).
Temporary (temp) basal rate
Allows you to temporarily change the current basal rate for a duration of 30 minutes up to 24 hours
(limited by Maximum basal setting). The temp basal rate can be set to either Percent of basal or Insulin
rate.
Time and date screen
12-hour or 24-hour formats. Pump users set the time/date, including the year, month and day. The date is
included in the STATUS screen. Time always appears on the top of the screen.
Environmental conditions
• Pump operating temperature range: from 37°F (3°C) to 104°F (40°C)
• Air pressure range: from 700 hPa to 1060 hPa (from 10.2 psi to 15.4 psi)
• Operating humidity range of the case: 20% to 95%. This requirement exceeds EN 60601-2-24:1998,
Section 10.2.1b (20% to 90%).
Radio frequency (RF) communication specifications
• Pump to transmitter frequency: 916.5 MHz; proprietary Medtronic protocol; range up to 6 feet
(1.8 meters)
• Pump to meter frequency: 916.5 MHz; proprietary Medtronic protocol; range up to 4 feet (1.2 meters)
• Pump to remote frequency: 916.5 MHz; proprietary Medtronic protocol; range up to 24 inches
(approximately 60 cm)