What to do
When an alarm is triggered, the pump goes into Attention mode and an
alarm message shows on the screen. The pump then defaults to the HOME screen. Do these steps when you
get an alarm:
1 View the alarm: From the HOME screen, press any button to see the alarm message.
2
Read all of the alarm text. There are instructions on how to fix the alarm condition. (Press
to
read more text, if available.)
3 Clear the alarm. Press ESC then ACT after you read the alarm instructions.
4 The HOME screen appears.
5 Follow the instructions that appeared with the alarm to fix the alarm condition.
6 Check your settings. Check your time, date, basal and other settings to make sure they are correct.
Alarm conditions
Alarms put the pump in Attention mode.
A (ALARM)
This alarm shows the letter A followed by two numbers. A-alarms cause all insulin delivery to stop. Your
pump settings are retained. If this alarm repeats often, call our 24 Hour HelpLine for assistance.
AUTO OFF
Alerts you that no buttons were pressed during the time limit you set for the AUTO OFF DURATION feature,
and so insulin delivery has been stopped.
BATT OUT LIMIT
Occurs if the battery has been out of the pump for more than five minutes. Verify that the pump time and
date are correct. If not correct, go to the UTILITIES MENU and reset the date and time.
Troubleshooting, alerts and alarms 165
Chapter 13