5 While the pump turns on, it will show one or more screens until
the HOME screen appears.
If the HOME screen does not appear, do these steps:
a. Check that the battery is inserted correctly. If the battery has
been installed backwards, remove the battery and install it
properly.
b. If your pump still does not turn on or you get a FAILED BATT TEST alarm, remove and replace the
battery with a new one.
c. If the pump is still not on, call our 24 Hour HelpLine.
6 Check to make sure the time and date are correct. If more than five minutes have passed since you
removed the battery, you will be prompted to check the time and date. Refer to the Setting the time
and date section in the Basic programming chapter for programming instructions.
7 Press ESC to view the STATUS screen, making sure no alarms are active. If an alarm is active, follow
the instructions on the screen.
Pump buttons
The buttons on the pump are used to navigate through the menus and screens, and to program the
features of the pump.