Meter option
About meters
You can set up your pump to automatically receive your blood
glucose readings from Bayer's CONTOUR
®
NEXT LINK Wireless
Meter, Bayer's CONTOUR
®
LINK Wireless Meter, a OneTouch
®
UltraLink
®
meter, or a BD Paradigm Link
®
glucose meter. Your
pump is set at the factory with the meter option turned off. See the Add, delete, review meter IDs section
in this chapter to turn the meter option on, and to enter your meter ID. Programming your meter ID links
your pump to the meter. Each meter has its own unique ID. You can link up to three meters to your pump.
If you do not link the meter to your pump, you must enter your blood glucose readings manually.
When the pump is idle (at the HOME screen), it will beep or vibrate when it receives a blood glucose
reading from the meter. The reading will appear on the pump screen. If your BG reading is not within your
set BG Target range, the pump beeps three times. For BG readings that are above the set BG Target
range, use the Bolus Wizard feature to enter a correction bolus.
NOTE: The use of RF (radio frequency) devices with the pump reduces pump battery life.
You have to turn on the meter option to add, delete or review the meter ID(s) programmed in your pump.
The meter ID is printed on the back of the meter. See the user guide that came with your meter for
detailed information on how to use it.
Meter rules
If you want your pump to communicate with the meter, the following conditions must apply:
1 The meter option must be turned on and programmed. Refer to the instructions in this section.
2 Your pump must be within 4 feet (1.2 meters) of your meter to receive the blood glucose reading.
3 The pump cannot have a LOW BATTERY alert condition.
Meter
BG readings
Meter option 147
Chapter 10